The use of cemented dual mobility acetabular cup in revision hip arthroplasty


Department of Orthopedic Surgery, Ain Shams University, Cairo, Egypt

Date of Submission24-Sep-2019
Date of Decision03-Oct-2019
Date of Acceptance14-Oct-2019
Date of Web Publication18-Aug-2020

Correspondence Address:
MD Ayman Fathy
Department of Orthopedic Surgery, Ain Shams University Hospitals, 2 Saad Abdelwares Street, New Nozha, Cairo, 11769
Egypt
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/eoj.eoj_45_19

Rights and Permissions
  Abstract 


Purpose Instability is always an important problem after both primary and revision total hip arthroplasty (rTHA). Dual mobility or tripolar unconstrained acetabular components give a solution for preventing and treating instability. Results of many studies using dual mobility cups (DMC) with mid-term to long-term follow-up support their effectiveness. Complications such as intraprosthetic dislocation (IPD) and accelerated wear have been studied, although they seem to be less significant in older, low-demand patients. However, their use in younger patients should be with caution owing to the lack of current data concerning this high-demand patient population. The aim of this study is to analyze the short-term results of cemented DMCs in rTHA. Specifically, we evaluated the following: clinical outcomes and scoring, dislocation, IPD rates, and other complications.
Patients and methods The results of a single design of DMC was prospectively evaluated in a continuous series of 20 rTHAs. Follow-up period was of at least 2 years. Mean age of the patients was 66.8 years old.
Results At the 2-year mean follow-up, the dislocation rate was 5%, and the IPD rate was 0%. Improvement of the hip score, function, pain, and range of motion was noticed.
Conclusion DMCs demonstrated a low dislocation rate in rTHA but did not solve problems related to perioperative technical errors. Moreover, IPD did not appear to be a concern when compared with the gain of preventing instability.

Keywords: dislocation, dual mobility, revision hip arthroplasty, tripolar


How to cite this article:
Fathy A, Ashoub M. The use of cemented dual mobility acetabular cup in revision hip arthroplasty. Egypt Orthop J 2019;54:168-73

How to cite this URL:
Fathy A, Ashoub M. The use of cemented dual mobility acetabular cup in revision hip arthroplasty. Egypt Orthop J [serial online] 2019 [cited 2020 Sep 23];54:168-73. Available from: http://www.eoj.eg.net/text.asp?2019/54/2/168/292411




  Introduction Top


Total hip arthroplasty (THA) is considered one of the most successful surgical procedures providing pain relief and improvement of quality of life in patients with advanced hip arthritis that is not improving with nonoperative treatments [1],[2]. As health services continue to improve and life expectancy increases, the need for total joint replacement will increase to cope with the basic needs of this growing portion of the population [2],[3].

Dislocation and repeated instability continue to be the most common indication for both revision and re-revision THA, with dislocation rates after primary THA range from 0.3 to 9% [4],[5],[6], whereas after revision total hip arthroplasty (rTHA) reportedly range from 5.1 to 27% of cases. Repeated hospital admission and revision surgeries increase economic cost as the surgical treatment of a dislocating THA can raise cost by 148% [1].

Over the past few years, large femoral heads, constrained prosthesis, and dual mobility acetabular cups have been introduced to solve this challenge with varying success [7],[8]. Although constrained acetabular cups showed high rates of mechanical loosening, dislocations, and limited of range-of-motion, large femoral heads showed also problems related to accelerated wear and increased incidence of local tissue reactions [9].

Recently, dual mobility (DM) prostheses were introduced with design concepts that address instability and avoid the problems of large femoral heads and constrained cups [10] In a DM system, a small-diameter femoral head articulates with a polyethylene liner and together form a large femoral head construct which in turn articulates, while still mobile within an inner acetabular shell [11]. Therefore, this design gives greater impingement-free range of motion and a larger head-to-neck ratio, resulting in increased jump distance, reducing the risk of intraprosthetic dislocation (IPD) [3],[12],[13],[14],[15].

This study reviews the use of these cups in rTHA, particularly where stability is in doubt. Dual mobility cups (DMC) may be used in risky primary total hip replacement, such as elderly patients with increased comorbidities and those with a neuromuscular diseases. They are used at revision surgery where the risk of dislocation is high, such as in patients with many prior dislocations, patients with aseptic loosening, or those with abductor deficiency [16].

The current series prospectively evaluated the outcome of a single-design DMC in rTHA with respect to clinical outcomes and scoring, dislocation, and IPD rates at short-term follow-up [4].


  Patients and methods Top


Between November 2013 and November 2016, 20 DMCs were implanted during cup exchange in revision hip arthroplasty in 20 patients and were prospectively included in our study and a follow-up of at least 2 years was done. Informed consent (approved by the Research Ethics Committee in Ain Shams University, Faculty of Medicine) was obtained from all individual participants included in the study. The reasons for choosing a DMC were older age and risk factors for hip instability (i.e. multiple revisions or revision for hip instability or old infection). The indications for cup exchange were for recurrent dislocation (n=3), aseptic loosening (n=16), and infection (n=1) with two-stage surgery done ([Table 1], [Table 2]).
Table 1 Indications for surgery for the patients included in the study

Click here to view
Table 2 Number of patients requiring graft for reconstruction of acetabular defect

Click here to view


Exclusion criteria were primary THA procedure, revision performed in case of bone tumors, young age group, and femoral-only revision.

The eight females and 12 males had a mean age of 66.8 years (range, 55–80 years) at revision surgery.

The clinical data, Harris hip score [17], and radiological findings were assessed and recorded preoperatively, immediately postoperatively, 6 weeks, 6 months, 1 year, and 2 years postoperatively. The radiographic evaluation was done with an anteroposterior view of the pelvis and anteroposterior and lateral views of the operated hip ([Figure 1]). Osteolysis and radiolucent lines in the acetabulum were evaluated on serial radiographs in the DeLee and Charnley zones. Loosening was defined as more than 3° of change in the cup inclination angle or more than 3 mm of cup migration [18].
Figure 1 Two-stage revision surgery with the use of cemented dual mobility acetabualar cup. (a) Preoperative radiography with acetabular and femoral septic loosening, (b) removal of the prosthesis and application of a spacer, and (c) application of cemented dual mobility cup after infection subsided.

Click here to view


All revisions were performed through the modified Hardinge approach by two senior surgeons. All patients were positioned in the dead lateral decubitus position. The preservation of the glutei, their nerve supply, and muscular attachment to the femur was done. We suspended the gluteus medius and gluteus minimus on stay sutures and re-sutured them again at the end of operation rather than detaching them from the bone. This preserves the gluteus minimus and avoids injury to the superior gluteal nerve.

After acetabular component removal and granulomatous tissue debridement, acetabular bone deficiency was addressed and graded according to the 4-grade classification of the AAOS [19].

After 4 min of manual mixing at controlled-operating room temperature (19–22°C) and humidity (43–53%), a doughy texture of the cement was obtained. Then, the cement was thickly applied into the shell. The DM (Avantage; Biomet, Warsaw, Indiana, USA) cup was placed into the construct using manual pressure and centralization with particular attention to ensure a 2- to 3-mm uniform thickness of the cement mantle around the DMC [20] ([Figure 2]).
Figure 2 Diagram explaining motions occurring in the dual mobility cup. The first motion occurring between the head and the polyethylene liner, and then the second motion between the polyethylene liner and the cemented cup [6].

Click here to view


After application of the femoral stem, assembly of metal head into mobile polyethylene liner was done ([Figure 3]) and introduced over the stem ([Figure 4]), and then reduction of the prosthesis was done.
Figure 3 Assembly of the metal head into the mobile polyethylene liner.

Click here to view
Figure 4 Head polyethylene liner introduced over the stem.

Click here to view


Postoperative physical therapy with full weight bearing was begun the day after surgery. For patients in whom complex acetabular and/or femoral reconstruction requiring bone graft was performed, partial weight bearing was prescribed during 6 weeks postoperatively and full weight bearing thereafter.


  Results Top


Follow-up in these 20 cases was at least 24 months.

Of the 20 assessable cases, one underwent debridement for repeated wound infection, and one case had dislocation of the joint between the cemented cup and the outer polyethylene, requiring reduction by external maneuvers under general anesthesia 11 months after cup exchange for hip instability. No patients experienced IPD.

Mean cup inclination in the coronal plane was 44.9° (38°–54°). A continuous 2-mm radiolucent line was visible in one asymptomatic patients. The line remained stable over time and was associated with osteolysis in zones 1.

There was no significant difference between acetabular-only revisions and both component revisions respectively (P=0.8). No recurrence of the dislocation was reported after appropriate management.

The final Harris hip scoring was much better with significant improvement of the overall scoring after the total period of follow-up with better quality of life ([Table 3]).
Table 3 Comparison of Harris hip scoring between preoperative, 6 weeks, 6 months, 1 year, and 2 years postoperatively

Click here to view



  Discussion Top


The outcome was satisfactory, with a single case of late dislocation, and is comparable to previously reported rates of DMC survival in rTHA [2],[18], but comparisons between studies are difficult because patients and indications vary greatly from one study to another.

While discussing our results, we have to keep in mind some points. Our goal was to assess DMC dislocation rates after revision surgery for any reason. Second, the mean follow-up period is 2-year follow-up, which may seem insufficient; however, it should be kept in mind that a substantial proportion of complications develop within the first 3 years.

Instability is among the main complications of prosthetic revision surgery. The dislocation rate in our study was one case which is comparable to the rates found in other studies of DMC use for rTHA [21],[22]. Garbuz et al. [23] reported benefits from using large-diameter heads (36 and 40 mm) to prevent instability in exchange THA, but they have the major drawback of inducing marked polyethylene wear. The risk of wear is lower with DMCs, as reported by Adam et al. [24].

Several studies demonstrated lower loosening rates with DMCs than with constrained cups [18]. The low dislocation rate supports our policy of using DMCs almost routinely for rTHA. However, the dislocation rate was higher than for primary THA [25], in which the dislocation rates after DMC implantation are less than 1%.

Numerous factors influence the stability of a THA, including implant positioning, the number of previous surgical procedures, and the presence of muscle damage related to extensive dissection. These risk factors for dislocation are more common in patients undergoing revision compared with primary THA.

Cemented metal cups are associated with high rates of loosening and polyethylene wear. In our study, one hip with a DMC cemented onto the bone was surrounded by a radiolucent line, indicating a need for close monitoring. Other studies found no cases of loosening of DMCs cemented onto bone [18], a fact that may reflect decreased stresses at the cement-bone interface.

The jumping distance represents the prosthetic stability of any hip arthroplasty, with inverse relation between the risk of dislocation and this distance. The jumping distance of a dual mobility construct is greater than that of any acetabular cup design. That is why, the dual mobility design is theoretically the most stable device. However, important rules must always be followed during implantation and orientation of a DMC [4].

DMCs have gained worldwide attention as an option in the prevention and treatment of instability particularly in rTHA. Indeed, several European studies reported that DMCs could offer the benefit of increased stability without compromising clinical outcomes and implant fixation durability [22].

Concerns such as IPD due to accelerated polyethylene wear have been raised in literature, although these complications are likely to be less significant in older and lower-activity level patients such as most of the patients undergoing rTHA [26].

In our study, we did not experience IPD, which was almost similar to previously reported rates of this specific complication, ranging from 0.28 to 1% at 10-year follow-up.


  Conclusion Top


Instability remains a significant issue after both primary and rTHA. DM or tripolar unconstrained acetabular components can provide a viable alternative in preventing and treating instability.

Our findings indicate that use of a cemented DMC reduces the short-term to mid-term risk of a second revision in first-time revisions compared with classic cup designs with low dislocation and complication rates. Longer follow-up is needed to establish any long-term clinical advantages when DMCs are used in revisions performed owing to dislocation [2],[3],[13].

In conclusion, DMCs demonstrated a low dislocation rate in rTHA. Importantly, our result emphasizes the ability of DMCs to reduce the risk of instability even in the cases of acetabular-only revisions. However, DMCs did not compensate for potential perioperative technical errors; optimal orientation of the DMC and restoration of the abductor mechanism and leg length should be achieved during revision. In addition, IPD did not appear to be a concern with respect to the benefit in term of instability prevention [4].

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
De Martino I, Triantafyllopoulos GK, Sculco PK, Sculco TP. Dual mobility cups in total hip arthroplasty. World J Orthop 2014; 5:180–187.  Back to cited text no. 1
    
2.
Mohaddes M, Cnudde P, Rolfson O, Wall A, Kärrholm J. Use of dual-mobility cup in revision hip arthroplasty reduces the risk for further dislocation: analysis of seven hundred and ninety one first-time revisions performed due to dislocation. Int Orthop 2017; 41:583–588.  Back to cited text no. 2
    
3.
Levin JM, Sultan AA, O’Donnell JA, Sodhi N, Khlopas A, Piuzzi NS, Mont MA. Modern dual-mobility cups in revision total hip arthroplasty: a systematic review and meta-analysis. J Arthroplasty 2018; 33:3793–3800.  Back to cited text no. 3
    
4.
Wegrzyn J, Tebaa E, Jacquel A, Carret JP, Bejui-Hugues J, Pibarot V. Can dual mobility cups prevent dislocation in all situations after revision total hip arthroplasty? J Arthroplasty 2015; 30:631–640.  Back to cited text no. 4
    
5.
Nho SJ, Kymes SM, Callaghan JJ, Felson DT. The burden of hip osteoarthritis in the United States: epidemiologic and economic considerations. J Am Acad Orthop Surg 2013; 21 (Suppl 1):S1–S6.  Back to cited text no. 5
    
6.
Van Heumen M, Heesterbeek PJ, Swierstra BA, Van Hellemondt GG, Goosen JH. Dual mobility acetabular component in revision total hip arthroplasty for persistent dislocation: no dislocations in 50 hips after1-5years. J Orthop Traumatol 2015; 16:15–20.  Back to cited text no. 6
    
7.
Plummer DR, Christy JM, Sporer SM, Paprosky WG, Della Valle CJ. Dual-mobility articulations for patients at high risk for dislocation. J Arthroplasty 2016; 31 (9 Suppl):131–135.  Back to cited text no. 7
    
8.
Caton JH, Prudhon JL, Ferreira A, Aslanian T, Verdier R. A comparative and retrospective study of three hundred and twenty primary Charnley type hip replacements with a minimum follow up of ten years to assess whether a dual mobility cup has a decreased dislocation risk. Int Orthop 2014; 38:1125–1129.  Back to cited text no. 8
    
9.
Massin P, Orain V, Philippot R, Farizon F, Fessy MH. Fixation failures of dual mobility cups: a mid-term study of 2601 hip replacements. Clin Orthop Relat Res 2012; 470:1932–1940.  Back to cited text no. 9
    
10.
Philippot R, Camilleri JP, Boyer B, Adam P, Farizon F. The use of a dual-articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow-up of 15 years. Int Orthop 2009; 33:927–932.  Back to cited text no. 10
    
11.
Leiber-Wackenheim F, Brunschweiler B, Ehlinger M, Gabrion A, Mertl P. Treatment of recurrent THR dislocation using of a cementless dual-mobility cup: a 59 cases series with a mean 8 years’ follow-up. Orthop Traumatol Surg Res 2011; 97:8–13.  Back to cited text no. 11
    
12.
Darrith B. Outcomes of dual mobility components in total hip arthroplasty: a systematic review of the literature. Bone Joint J 2018; 100-B:11–19.  Back to cited text no. 12
    
13.
Simian E. Dual mobility cup in revision total hip arthroplasty: dislocation rate and survival after 5 years. Orthop Traumatol Surg Res 2015; 101:577–581.  Back to cited text no. 13
    
14.
Guyen O, Chen QS, Bejui-Hugues J, Berry DJ, An KN. Unconstrained tripolar hip implants: effect on hip stability. Clin Orthop Relat Res 2007; 455:202–208.  Back to cited text no. 14
    
15.
Guyen O, Pibarot V, Vaz G, Chevillotte C, Béjui-Hugues J. Use of a dual mobility socket to manage total hip arthroplasty instability. Clin Orthop Relat Res 2009; 467:465–472.  Back to cited text no. 15
    
16.
Ko LM, Hozack WJ. Acetabular revision.The dual mobility cup what problems does it solve? Bone Joint J 2016; 98-B:60–63.  Back to cited text no. 16
    
17.
Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 1969; 51:737–755.  Back to cited text no. 17
    
18.
Hamadouche M, Biau DJ, Huten D, Musset T, Gaucher F. The use of a cemented dual mobility socket to treat recurrent dislocation. Clin Orthop Relat Res 2010; 468:3248–3254.  Back to cited text no. 18
    
19.
D’Antonio JA, Capello WN, Borden LS, Bargar WL, Bierbaum BF, Boettcher WG et al. Classification and management of acetabular abnormalities in total hip arthroplasty. Clin Orthop Relat Res 1989; 243:126–137.  Back to cited text no. 19
    
20.
Wegrzyn J, Thoreson AR, Guyen O, Wegrzyn J, Thoreson AR, Guyen O et al. Cementation of a dual-mobility acetabular component into a well-fixed metal shell during revision total hip arthroplasty: a biomechanical validation. J Orthop Res 2013; 31:991.  Back to cited text no. 20
    
21.
Schneider L, Philippot R, Boyer B, Farizon F. Revision total hip arthroplasty using a reconstruction cage device and a cemented dual mobility cup. Orthop Traumatol Surg Res 2011; 97:807–813.  Back to cited text no. 21
    
22.
Hailer NP, Weiss RJ, Stark A, Kärrholm J. Dual-mobility cups for revision due to instability are associated with a low rate of re-revisions due to dislocations: 228 patients from the Swedish Hip Arthroplasty Register. Acta Orthop 2012; 83:566–571.  Back to cited text no. 22
    
23.
Garbuz DS, Masri BA, Duncan CP, Greidanus NV, Bohm ER, Petrak MJ et al. The Frank Stinchfield Award: dislocation in revision THA: do large heads (36 and 40 mm) result in reduced dislocation rates in a randomized clinical trial? Clin Orthop Relat Res 2012; 470:351–356.  Back to cited text no. 23
    
24.
Adam P, Farizon F, Fessy MH. Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants. Rev Chir Orthop 2005; 91:627–636.  Back to cited text no. 24
    
25.
Vielpeau C, Lebel B, Ardouin L, Burdin G, Lautridou C. The dual mobility socket concept: experience with 668 cases. Int Orthop 2011; 35:225–230.  Back to cited text no. 25
    
26.
Vasukutty NL, Middleton RG, Matthews EC, Young PS, Uzoigwe CE, Minhas TH. The double-mobility acetabular component in revision total hip replacement.The United Kingdom experience. J Bone Joint Surg Br 2012; 94:603.  Back to cited text no. 26
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Patients and methods
Results
Discussion
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed252    
    Printed4    
    Emailed0    
    PDF Downloaded22    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]